GDPD-414H Handheld partialis officii Detector
●Portable ABS engineering gb.
●2 vel 4-canalis synchrona notitia sui iuris acquisitio.
●8.1inch 1280*800 IPS tactus screen.
●Utens rubris, flavis, et caeruleis coloribus ad severitatem partialis missionis indicandam.
●Mediocris tempus molestum est super 50000hours.
●Pondus unitatis principale minus est quam 0.8kg.
●ARMARIUM in ratione systematis embedded in Analyse software fundatur.Propono programmata secundum fenestras systematis.
PD signum acquisitio exercitum | |
CPU Opus frequentia | 800MHz |
Sistema operatum | Linx embedded operating ratio |
Wired network portum | LAN network portum |
Wireless network portum | Inaedificata wireless WiFi |
Ratio currit memoria | 512M |
Ratio repono memoriae | 256M |
Data frequentia acquisition | 80MHz |
Ultrasonic deprehensio channel | |
Mensurae range | 0-60mV |
Frequency deprehendatur range | 20~200kHz |
UHF deprehendatur channel | |
Deprehensio frequency | 300~500MHz |
Mensurae range | -80~10dBm |
Error | ±1dBm |
Resolution | 1dBm |
HFCT deprehendatur channel | |
Frequency range | 0.5~100MHz |
Error | ±1dB |
Dynamic Range | 60dB |
Mensurae range | 0-100mV |
Sagacitas | 1dB |
TEV deprehendatur channel | |
Frequency range | 3~100MHz |
Error | ±1dB/mV |
Sensibilitas | 0.01mV |
Mensurae range | 0-60dB/mV |
Resolution | 1dBm/mV |
Pugna | |
Inaedificata altilium | Lithium altilium, 12V, 2000mAh |
utere tempore | circa VI horas |
dato tempore | Circa II horas |
altilium praesidium | Plus-voltage and over-current protection |
altilium præcipiens | |
Rated voltage | 12.6V |
Incurrentes current output | 2A |
Operating Temperature | -20℃-60℃ |
humiditas operating | <80% |
Handheld propono terminatio (industrialis gradus) | |
cpu | Intel Quad Core Atom Z3735F |
GPU | Intel HD Graphic (Gen7) |
Flash | 32GB |
aries | 2GB |
Sistema operatum | Windows10 |
Ostendere | 8.1 inch 1280×800 IPS screen |
Network interface | WiFi et Bluetooth |
Pugna | 3.7V 8500mAH polymer lithium ion altilium |
Magnitudo | |
PD acquisition exercitum magnitudine | 240mm* 165mm*55mm |
Acquisitio exercitus pondus PD | 0.65kg |
Ostentationem tabella terminali magnitudine | 395mm* 295mm* 105mm |
Ostentationem tabella terminali pondere | 0.85kg |
Arca altiore magnitudine | 570mm* 360mm* 240mm |
Situs laboris | |
opus temperatus | -20℃~50℃ |
Environment humiditas | 0~90%RH |
IP level | 54 |